B-337 A Rapid and Selective LC-MS/MS Workflow for Methylmalonic Acid Using Strata™ SE SLE
TL;DR
The developed streamlined, selective, and high throughput LC MS/MS method for MMA quantification in human serum exhibited excellent linearity and strong analytical performance at low concentrations, supporting its utility for routine clinical research applications focused on evaluating functional vitamin B12 status in biological fluids.
Abstract
Methylmalonic acid (MMA) is a sensitive biomarker of functional vitamin B12 status and is often measured in clinical research due to its role in cellular metabolism and its association with neurological and cardiovascular dysfunction. Accurate quantification of MMA in human serum is challenging because of low endogenous concentrations, high polarity, and structural similarity to isobaric compounds such as succinic acid, requiring a highly selective LC separation and sample extraction method for robust LC MS/MS analysis. This work aimed to develop a streamlined, selective LC MS/MS method for MMA from human serum using a thermally modified polar embedded C18 column possessing a positively charged surface, and Strata™ SE supported liquid extraction (SLE), optimized to improve analyte recovery, reduce matrix effects, and enhance throughput in clinical research workflows. A 200 µL aliquot of human serum was combined with an equal volume of 4.5 M aqueous formic acid and applied to a Strata™ SE SLE MAX 96 well plate. After a brief vacuum pulse and 5 minute equilibration to ensure complete sample absorption onto the SLE sorbent, MMA was extracted using two 900 µL aliquots of ethyl acetate/hexane (9:1, v/v). Organic eluates were evaporated to dryness under nitrogen and reconstituted in a mobile phase compatible solvent. Chromatographic separation was achieved on a Luna Omega PS C18 column (3.0 × 150 mm, 3 µm). The mobile phase consisted of 0.5% formic acid in water (A) and acetonitrile (B), using a linear gradient from 85% to 55% B at 0.5 mL/min. Total run time, including re equilibration, was 5 minutes. Detection was performed on a SCIEX 6500 triple quadrupole mass spectrometer in negative electrospray ionization mode. The positively charged polar embedded stationary phase provided distinct HILIC like retention and enabled baseline separation of MMA from succinic acid, with a retention time difference of 0.8 minutes. This selectivity reduced matrix-derived interferences and produced a cleaner chromatographic background relative to conventional reversed phase columns. Calibration using extracted human serum demonstrated excellent linearity across a 1000 fold concentration range (1–1000 ng/mL; r² = 0.998, 1/x weighting). At the LOD, recovery was 103%, with process efficiency and matrix effect of 79% and 80%, respectively, indicating robust analyte retention and minimal ion suppression. This study describes a sensitive, selective, and high throughput LC MS/MS method for MMA quantification in human serum. The integration of Strata™ SE SLE provided efficient matrix removal and high analyte recovery, while the Luna Omega PS C18 column operated under HILIC conditions delivered the chromatographic selectivity required to fully resolve MMA from succinic acid and other matrix interferences. The method exhibited excellent linearity and strong analytical performance at low concentrations, supporting its utility for routine clinical research applications focused on evaluating functional vitamin B12 status in biological fluids.