Oct 2026· Biomedical engineering and physics express· Vol 12, pp. 055073· 0 citations· 31 references
PhysicsMedicine
TL;DR
This work presents the first experimental evaluation of the easyPET, a compact and cost-effective preclinical positron emission tomography scanner, as a potential tool for quality control assessment and validate the potential of easyPET.3D as a promising, accessible solution for proton range verification.
Abstract
Accurate range verification is a critical component of proton therapy, ensuring that the high dose gradients characteristic of Bragg peaks are delivered precisely to the target volume. This work presents the first experimental evaluation of the easyPET.3D, a compact and cost-effective preclinical positron emission tomography scanner, as a potential tool for quality control assessment. Homogeneous PMMA phantoms were irradiated at the Paul Scherrer Institute Gantry 1 facility using proton beam energies ranging from 138 MeV–195 MeV. The resulting 3D β+ activity distributions were reconstructed and analyzed to assess the system’s spatial accuracy, stability, and sensitivity. For the intermediate energy of 160 MeV, the measured activity profile showed a high degree of agreement with Monte Carlo (GATE) simulations, with a difference of 0.24 mm in the R50% position. The system exhibited robust performance under varying irradiation conditions. Doubling the number of delivered protons relative to a clinically relevant value resulted in a deviation of only 0.4 mm in the calculated range, confirming the method’s reproducibility. Furthermore, the system successfully resolved distinct beam paths in a dual-energy irradiation mode, distinguishing tracks separated vertically by 10 mm and quantifying longitudinal range differences. Sensitivity tests indicated that the system is capable of detecting longitudinal phantom displacements as small as 1 mm. The relationship between expected and measured displacements followed a linear trend with a slope of 1.13, indicating a systematic overestimation of 13%, and an offset of 0.45 mm, reflecting a small baseline bias. While further optimization of detector sensitivity and reconstruction algorithms is required to reduce acquisition times for routine clinical workflows, these findings validate the potential of easyPET.3D as a promising, accessible solution for proton range verification.
This Monte Carlo study investigated a dual-head SPECT-based system for prompt-gamma (PG) range verification in proton therapy, focusing on energy-window and collimator septal thickness effects on PG emission profiles and distal dose fall-off. Using GATE, correlations between PG emissions (4.44, 5.24, 6.13 MeV) and prot...
Sharareh Gharib, P. Taherparvar· Biomedical engineering and p...· 0 citations
Treatment-planning studies indicate potential dosimetric advantages of very high-energy electron (VHEE) radiotherapy for deep-seated tumours, but the dose models on which these studies rely have few experimental benchmarks above 100 MeV. Here, we report radiochromic film dosimetry of 150 and 200 MeV beams at CERN’s C...
Fabio S. D'Andrea, Adam Aitkenhead, R. Chuter et al.· Communications Physics· 0 citations
While carbon ions provide better RSP reconstruction precision essential for complex treatment plans in heterogeneous anatomy, helium and proton imaging offer exceptional dose sparing, making them particularly advantageous for pediatric patients and frequent adaptive replanning scenarios.
Z. Jólesz, G. Bíró, G. Papp et al.· Physics in Medicine and Biol...· 0 citations
Helical Tomotherapy modulated the radiation beams intensity using binary multileaf movement synchronization with couch translation and gantry rotation. This complexity of delivery required a valid patient-specific quality assurance (PSQA) system to ensure the accurate delivery of treatment plans. The Delta 4 plus is th...
Objective. Linear energy transfer (LET)-informed proton therapy has the potential to improve biological treatment planning beyond the conventional assumption of a constant relative biological effectiveness. Its clinical translation, however, requires experimental tools capable of verifying spatially resolved dose and d...
M. L. Jensen, M. van Heerden, L. Stolarczyk et al.· Physics in Medicine and Biol...· 0 citations
Results indicate that the detector can provide reliable and reproducible data and demonstrate the applicability of active-matrix technology to experimental 3D dosimetry in radiotherapy.
D. Roa, M. Risco-Castillo, R. Romero et al.· Journal of Physics, Conferen...· 0 citations