Aug 2026· Optical Engineering: The Journal of SPIE· Vol 65, pp. 084102 - 084102· 0 citations· 25 references
Engineering
TL;DR
The accuracy and repeatability of the proposed mechanically scanning approach under model eye conditions are demonstrated and support its further development as a platform for future peripheral aberration studies.
Abstract
Abstract. Peripheral refraction is thought to play an important role in the onset and progression of myopia, highlighting the need for reliable methods to quantify peripheral optical aberrations in the eye. We present a mechanically scanning Hartmann–Shack system and validates it using a model eye. The system coordinates the mirror rotation and translation with synchronous motion of the wavefront sensor to perform peripheral aberration measurements. For a 6-mm entrance pupil, central field refractive errors from −6 to +6 D were measured, and horizontal meridian peripheral aberrations were acquired over a 0- to 30-deg field of view in 2.5-deg steps. The experimental measurements were compared with corresponding Zemax-based simulations. Five repeated measurements showed good agreement with the simulated central field defocus, with a linear fit slope close to 1. In off-axis fields, the experimental and simulated results demonstrated reliable measurement performance, with spherical equivalent differences below 0.10 D and a maximum astigmatic component J0 difference of 0.19 D. These results demonstrate the accuracy and repeatability of the proposed mechanically scanning approach under model eye conditions and support its further development as a platform for future peripheral aberration studies.
PURPOSE
To compare the optical performance of the Clareon PanOptix and Clareon PanOptix Pro intraocular lenses (IOLs) using interferometric metrology and identify measurable optical differences.
SETTING
Physiological Optics Laboratory, Daegu, Republic of Korea.
DESIGN
Experimental optical-bench investigation.
MET...
E. Jeong, S. Bang, Seth M. Pantanelli et al.· Journal of cataract and refr...· 0 citations
Chromatic confocal studies often report optical design and metrology performance using non-equivalent metrics, limiting reproducibility and practical comparison. We present a design-to-validation study built around a reproducible four-element, three-group all-spherical objective using CDGM catalog glasses. The objectiv...
By providing precise postoperative wavefronts and through-focus visual quality metrics, this approach offers a robust tool for clinical evaluation of patients implanted with these IOLs, and overcomes the phase asymmetry of segmented bifocal IOLs.
Suman Sangiri, Vyas Akondi· Journal of refractive surger...· 0 citations
Back vertex power is a fundamental parameter for evaluating the optical performance of contact lenses, and the accuracy of its measurement is directly related to the corrective efficacy of the lens. Traditional methods, such as lensometers, require physical contact and are prone to errors caused by contact pressure, le...
Jie Zhao, Yong-Jie Luo, Yao Tang et al.· Global Intelligent Industry...· 0 citations
Precise measurement of large mirrors is essential for high-performance space-based optics. Traditionally, laser interferometry provides high accuracy but faces challenges related to the requirements of computer-generated holograms (CGHs) and the transformation of standard wavefronts to match complex test surfaces. This...
Juliana M. Y. Tam, Henry Wong, Yixiao Han et al.· Metrology· 0 citations
The segmented primary mirror represents a major breakthrough in large-scale optical engineering, overcoming the long-standing difficulties associated with the fabrication, transportation, and installation of large monolithic primary mirrors. During manufacturing and alignment, interferometric testing based on computer-...