B-410 Assay Development and Validation for Therapeutic Drug Monitoring (TDM) of Treosulfan on LC-MS/MS in Human Plasma
TL;DR
A robust and sensitive HPLC-MS/MS method for quantification of treosulfan in human plasma was developed and validated, showing excellent performance with good linearity, low variability, high accuracy, minimal carry-over and matrix effects, as well as favorable stability under certain conditions.
Abstract
Treosulfan is a water-soluble bifunctional alkylating prodrug and structural analog of busulfan used as an important conditioning agent prior to hematopoietic stem cell transplantation (HSCT) in malignant and non-malignant diseases. Accurate testing of treosulfan concentrations in blood plasma is crucial for investigating the pharmacokinetic profile of treosulfan and thus for optimal treosulfan dosing in conditioning prior to HSCT, especially in pediatrics. We aimed to pioneer a robust and sensitive HPLC-MS/MS method for quantification of treosulfan in human plasma. Concentrations for treosulfan and its primary biologically active metabolite (2S,3S)-1,2-epoxy-3,4-butanediol 4-methanesulfonate (S,S-EBDM) were determined with two individual methods with different extraction and analysis. The analytes were extracted from 50 uL of dilute human plasma using protein precipitation with methanol. The resulting sample is analyzed on an Agilent Triple Quadrupole LC-MS 6470 equipped with a ZORBAX Eclipse Plus C18 RRHD 3.0 x 50mm 1.8-micron column and a 3.0 x 5mm guard column of the same packing. A 2.8-minute LC gradient method starting at 95% mobile phase A (2mM ammonium acetate in water with 0.1% formic acid, pH between 2 and 3) and 5% mobile phase B (2mM ammonium acetate in methanol with 0.1% formic acid, pH between 4 and 5) was applied with a flow rate of 0.5 mL/min. The injection volume is 2.0 µL. Qualifier and quantifier transitions from treosulfan and treosulfan-D4 internal standard were used to verify the identity and concentration. Method validation included accuracy, precision, calibration verification, sensitivity, specificity, carryover, matrix effects, dilution integrity, recovery, sample stability, and method comparison. Validation is performed in accordance with CLSI C62-Ed2, FDA Bioanalytical Method Validation, EMA Bioanalytical Method Validation, and Scientific Working Group for Forensic Toxicology (SWGTOX) Standard Practices for Method Validation in Forensic Toxicology guidelines. The analytical measurement range (AMR) for treosulfan was established at 500 to 250,000 ng/mL. With 5x dilution, the clinical reportable range (CRR) is 500 to 1,250,000 ng/mL. The assay showed excellent system suitability, linearity (R2 = 0.996), accuracy (recoveries 94%–108%), precision (within-day and between CV% were =3.6%), specificity (interference from the matrix, internal standard, and commonly used drugs were all = 20% of LLOQ), carryover (less than 15% of LLOQ) and sensitivity with LOD (26 ng/mL) and LOQ (500 ng/mL). A robust HPLC-MS/MS method for quantification of treosulfan in human plasma was developed and validated, showing excellent performance with good linearity, low variability, high accuracy, minimal carry-over and matrix effects, as well as favorable stability under certain conditions. This method can be used for clinical pharmacokinetics and real-time, targeted optimization of treosulfan dosing in pediatric and adult patients.