Skip to content

Dosimetric robustness of tumour- and fiducial-based tracking in Radixact lung stereotactic body radiotherapy: a four-dimensional CT-based simulation using imaging-derived tracking-error scenarios.

Jul 2026 · British Journal of Radiology · 0 citations
Medicine

TL;DR

For lung SBRT delivered with Radixact Synchrony, LwR was more robust than FwR in this cohort, for which all tumour-fiducial distances exceeded 20 mm, and patient-specific evaluation of 5-7 mm margins within OAR dose constraints may be necessary.

Abstract

Objectives

To compare the dosimetry and robustness of two Radixact Synchrony lung stereotactic body radiotherapy (SBRT) tracking strategies: tumour-based lung tracking with respiratory modelling (LwR) and fiducial-based tracking with respiratory modelling (FwR).

Methods

Ten stage I non-small-cell lung cancer patients were retrospectively analysed. SBRT plans (42 Gy in 4 fractions) were generated on maximum-exhalation computed tomography (CT) using planning target volumes (PTVs) created by 3-7 mm isotropic gross tumour volume (GTV) expansion. Each plan was mapped to all four-dimensional CT (4DCT) phases via isocentre shifts from phase-specific GTV (LwR) or individual fiducial displacement (FwR). Deformable dose accumulation yielded LwR, nearest FwR, and farthest FwR simulated plans. GTV/PTV V100%, representative organ-at-risk (OAR) doses, and robustness versus target localisation error, tumour-fiducial distance, and tumour volume were evaluated.

Results

With a 3-mm PTV margin, FwR showed larger losses in GTV coverage and PTV robustness than LwR (maximum decreases of 9.0% and 26.0%). FwR was less robust, particularly with unfavourable tumour-fiducial geometry. Larger PTV margins progressively mitigated FwR coverage degradation while maintaining acceptable OAR doses.

Conclusion

For lung SBRT delivered with Radixact Synchrony, LwR was more robust than FwR in this cohort, for which all tumour-fiducial distances exceeded 20 mm. When FwR is required, particularly under suboptimal tumour-fiducial geometry, a 3-mm margin should be used cautiously, and patient-specific evaluation of 5-7 mm margins within OAR dose constraints may be necessary. ADVANCES IN KNOWLEDGE A 4DCT-based simulation using imaging-derived tracking-error scenarios reveals dosimetric robustness differences between LwR and FwR tracking.

View source

Similar papers

Open access Sep 2026

Dosimetric Impact of Residual Intrafraction Motion in Gated MR-Guided Prostate SBRT With Focal Dose Intensification

Residual intrafraction motion produced directionally asymmetric coverage loss, with the largest effect on the boosted GTV, and motion-inclusive dose reconstruction enabled derivation of practical asymmetric margins that improved robustness while identifying the tradeoff between target coverage and delivery efficiency.

Ka-Ho Tam, S. Jogi, A. Sobremonte et al. · 0 citations
Open access Aug 2026

Dosimetric Evaluation of MR-Guided IMRT and VMAT Techniques for Early-Stage Glottic Cancer: A Pilot Comparative Planning Study

MR-guided IMRT provided superior dose conformity while maintaining target coverage comparable to TrueBeam STx with clinically acceptable OAR sparing, although selected laryngeal structures and the esophagus differed significantly.

V. Rudžianskas, E. Jaselskė, J. Laurikaitienė et al. · 0 citations
Open access Aug 2026

Evaluation of EPID-Based Transmission and DLG Correction Methods for Dosimetric Verification in HyperArc Single-Isocenter Multiple Target Radiosurgery

Background/Objectives: Accurate verification of single-isocenter multiple-target (SIMT) stereotactic radiosurgery is challenging owing to the complexity of multi-lesion delivery and the sensitivity of stereotactic dose gradients. We aimed to evaluate the efficacy of electronic portal imaging device (EPID)-based multile...

S. Oh, Sung Yeop Kim, J. Park et al. · 0 citations
Open access Aug 2026

Workflow and dosimetric results for direct-to-unit single-fraction adaptive lung SBRT delivered on the ONE-STOP prospective pilot clinical trial

CBCT-guided DTU single-fraction lung SBRT is feasible and dosimetrically similar to conventional workflows, enabling high-quality treatment in a single clinical visit, and is motivated by expanded application of DTU workflows within the clinic and expansion to other anatomical sites to reduce time toxicity and improve...

F. Forghani, T. Mazur, Joshua P. Schiff et al. · 0 citations

CBCT-based dose tracking in head and neck proton therapy: analysis of dose deviation for adaptive re-simulation assessment

Proton therapy has been increasingly utilized in head and neck cancer (HNC) patients due to its unique property of the Bragg Peak. However, proton dose delivery is highly sensitive to positional variations and anatomical changes occurring during the treatment course. These changes can lead to deviations between planned...

Nay Chi Tun Aung · 0 citations
Open access Aug 2026

Decoupling the dosimetric impact of setup errors and anatomical deformations in image-guided radiotherapy using diagnostic-quality FBCT and glmmLasso regularization

This study uses artifact-free fan-beam CT (FBCT) and a generalized linear mixed Lasso model to decouple interfraction setup errors and anatomical deformations within a simulated image-guided radiotherapy workflow, providing quantitative support for future site-specific ART research.

Lianzi Zhao, Ying Guo, Yuanshuai Di et al. · 0 citations