Skip to content
Open access

Characterisation and management of therapy-interrupting system errors associated with the use of the HomeChoice cycler for automated peritoneal dialysis

Aug 2026 · BMC Health Services Research · Vol 26 · 0 citations · 34 references
Medicine

TL;DR

The nature of therapy interrupting system errors in automated peritoneal dialysis (APD) cyclers of a German PD reference care centre is analysed and resolution strategies to these errors based on routinely collected health data and user experience are suggested to safely guide nephrology professionals and patients during troubleshooting.

Abstract

Methodological and technical education of nurses is required to address challenges of technology integration in peritoneal dialysis (PD). We analysed the nature of therapy interrupting system errors in automated peritoneal dialysis (APD) cyclers of a German PD reference care centre and provide applicable resolution proposals. We documented therapy interrupting system errors of HomeChoice Pro and HomeChoice Claria cyclers (Baxter, BX, Unterschleißheim, Germany) including the years 2015-2023. We systematically characterise and suggest resolution strategies to these errors based on routinely collected health data and user experience to safely guide nephrology professionals and patients during troubleshooting. The ‘reporting of studies conducted using observational routinely collected health data’ (RECORD)-statement was considered. We identified 49 different therapy interrupting system errors which were assigned to eight major categories: Errors associated to the pressure chamber (N=18), the Digitalboard/EEPROM (N=18, Electrically Erasable Programmable Read Only Memory), air in set (N=3), opening of the set’s door (N=3), those associated to the energy system (N=3), the heater system (N=3) and a variety of errors resulting in an abrupt termination or non-commencement of the treatment session without clear group assignment (undefined, N=1) or without sufficient documentation. In total, 138 error incidents were documented during the observational timeframe. In frequential order errors were attributed to the pressure chamber (21.74%,), and air in set (19.57%) where specifically error 2240 occurred in 17.39% of total cases, the Digitalboard/EEPROM (17.39%), the energy or battery (16.67%; No. 1032 in 19/23 cases), the set’s door (7.25%) and the heater (4.35%). In 17 cases (12.32%) no error number was documented by the patient or staff. One error cause was registered as undefined (0.72%). Overall, errors were rarely occurring in <0.25% of treatment sessions performed in the observational timeframe. These findings may be useful to guide training purposes for nephrology personnel and persons doing APD. For the manufacturer, the information provided may be helpful to improve error management and information transmission to the persons doing APD with the HomeChoice devices.

Read PDF

Similar papers

Review Open access Aug 2026

Approach to the prevention of medication errors using electronic barriers in PICU and NICU: an ESPNIC survey

Medication errors (MEs) are a major cause of preventable harm in critically ill pediatric patients. Electronic safety systems — such as Computerized Physician Order Entry (CPOE), smart pumps, and barcode medication administration — are increasingly adopted to reduce MEs, yet their implementation in pediatric and neonat...

A. Tessari, S. D. de Wildt, E. Buccione et al. · 0 citations
Open access Sep 2026

From manual to digital: prospective ICU evaluation of the FIZE kUO® for continuous urine output measurement

ICU nurses manage complex, time-sensitive care while also performing frequent manual tasks. Manual urine output (UO) measurement is one such task that remains labor-intensive, interruptive to workflow, and prone to delays - despite the need for timely and accurate recording. Automation may reduce this burden and improv...

Olivier Zerbib, Tawfik Mahamid, Noa Ben-Noon et al. · 0 citations
Aug 2026

Human factors testing in a novel automated peritoneal dialysis system.

The Archimedes APD system usability was found to be safe, accessible and easy for its intended users and use environments after the relatively short training period compared to the status quo.

Nupur Gupta, Vikram Aggarwal, James A. Sloand et al. · 0 citations
Review Open access Sep 2026

From Remote Monitoring to Actionable Care: Real-World Implementation of an Interoperable Digital Ecosystem for Peritoneal Dialysis

Background/Objectives: Remote patient monitoring (RPM) has expanded the capacity of peritoneal dialysis (PD) programs to supervise home-based therapy; however, its full clinical value may be limited when patient-generated data remain disconnected from electronic medical records (EMRs). This study evaluated the implemen...

A. Rivera, J. Vesga, M. Sanabria et al. · 0 citations
Review Open access Sep 2026

Standardizing the digital frontier: A Nordic survey of remote patient management practice patterns and variations in peritoneal dialysis care.

BackgroundRemote patient management (RPM) improves automated peritoneal dialysis (APD) care safety and efficiency, yet the detailed clinical pathway of its utilization remains variable. We investigated digital practice patterns, healthcare professional (HCP) perceptions, and potential gaps for RPMAPD implementation acr...

L. Dai, Giedre Martus, J. Kampmann et al. · 0 citations
Aug 2026

Development and Implementation of the MED4TUBE Evidence-Based Care Bundle to Reduce Medication Errors in Enteral Feeding Tubes: A Quality Improvement Study.

BACKGROUND Medication administration via enteral feeding tubes is a complex process and is often associated with preventable errors that may affect treatment effectiveness and patient safety. OBJECTIVE This study aimed to develop an evidence-based care bundle and evaluate its impact on medication errors in hospitaliz...

Sevgin Memili, R. E. Sezer, Kezban Akçay et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.