Interobserver Variability in the Identification of Keratoconus Suspects Among Teenagers and Young Adults: Comparison With Scheimpflug Tomographic Indices.

Aug 2026 · Cornea · 0 citations · 23 references
Medicine

TL;DR

Moderate interobserver agreement highlights the diagnostic uncertainty of identifying early keratoconus-suspect eyes and may contribute to variability in reported keratoconus prevalence.

Abstract

Purpose

To evaluate interobserver agreement between 2 experienced corneal specialists in classifying eyes as normal (N), keratoconus-suspect (KCN-S), or keratoconic (KCN) using Scheimpflug tomography, and to assess the discriminatory performance of tomographic asymmetry indices.

Methods

A total of 206 eyes from 103 participants (15-25 years) underwent Scheimpflug corneal tomography. Two experienced corneal specialists independently classified each eye based solely on four-map displays, without access to quantitative tomographic indices (BAD-D, ISV, IHD, IVA). Interobserver agreement was assessed using Cohen κ statistic. Discriminatory performance was evaluated by receiver operating characteristic (ROC) analysis and area under the curve (AUC), with observer classifications retrospectively compared with BAD-D, ISV, IHD, and IVA.

Results

Interobserver agreement was moderate (κ = 0.40, P <0.01). Agreement between observer classifications and BAD-D was substantial for observer 1 (κ = 0.70, P <0.01) and fair for observer 2 (κ = 0.35, P <0.01). Among the evaluated indices, BAD-D demonstrated the highest discriminatory performance (AUC = 0.94 and 0.85 using observer 1 and observer 2 classifications, respectively). In expert consensus cases, BAD-D achieved an AUC of 0.97. A BAD-D threshold of 1.10 yielded 90% sensitivity and 90% specificity.

Conclusions

Moderate interobserver agreement highlights the diagnostic uncertainty of identifying early keratoconus-suspect eyes and may contribute to variability in reported keratoconus prevalence. Among the evaluated tomographic indices, BAD-D showed the strongest agreement with expert classifications and the greatest discriminatory ability, supporting its use as an adjunctive tool for early screening in young individuals.

View source

Similar papers

Aug 2026

Corneal Densitometry Maps for Diagnosis of Common Corneal Diseases: A Masked Four-Reader Pilot Study.

PURPOSE To evaluate whether qualitative corneal optical densitometry (COD) maps alone allow diagnostic discrimination of common corneal diseases in a masked multireader setting. METHODS In this retrospective diagnostic accuracy study, 500 COD maps obtained with Scheimpflug imaging were retrospectively identified and...

E. Golschan, Tyll Jandewerth, K. P. Kaiser et al. · 0 citations
Open access Sep 2026

Measurement Repeatability and Concordance of Corneal Tomography in Keratoconus Using Anterion Versus Pentacam

Purpose: This study aimed to compare the repeatability and agreement of corneal tomographic measurements between the Anterion Cornea App and Pentacam HR in keratoconus (KCN). Methods: Eyes with KCN were imaged three consecutive times with each device. Intra-device repeatability was assessed using intraclass correlation...

Ramadhan Ahmed, K. Reddy, Mahad Mohamed et al. · 0 citations
Sep 2026

Quantitative evaluation of choroidal thickness and vascularity in subclinical keratoconus: a cross-sectional EDI-OCT study

The findings demonstrate increased CT in both CK and SK, whereas most quantitative choroidal vascularity parameters remained unchanged, suggesting that choroidal thickening may represent an early structural alteration in the keratoconus spectrum.

Hasan Kızıltoprak, Burak Ören, D. Kocabas et al. · 0 citations
Sep 2026

Tomographic and densitometric corneal alterations in asymptomatic offspring of patients with keratoconus: a Turkish cross-sectional study

Pentacam HR-derived tomographic and densitometric parameters may facilitate early identification of individuals with increased ectatic susceptibility and may facilitate early identification of individuals with increased ectatic susceptibility.

Halil İbrahim Ateşoğlu, Mine Turkay Atesoglu, Hasan Kızıltoprak et al. · 0 citations
Open access Aug 2026

Agreement of keratometry measurements for toric IOL planning in routine clinical practice: KR-800, IOLMaster 500, and Pentacam HR

The three devices showed acceptable agreement at the population level, and notably, the average magnitude of disagreement was limited, at the individual level, however, differences exceeded clinically relevant thresholds in a meaningful proportion of patients, including some toric IOL candidates.

Yeoeun Lim, Moonwon Hwang · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.