SAR validation of RF coils and implants in a B0‐free RF safety laboratory

Aug 2026 · Medical Physics (Lancaster) · Vol 53 · 1 citation · 49 references
Medicine

TL;DR

Direct SAR measurement provides a reliable and quantitative approach for validating RF coil and implant safety, matching the accuracy of conventional scanner‐based thermometry and simulation methods.

Abstract

As MRI continues to advance toward higher field strengths, RF safety assessment has become more complex. Accurate estimation of specific absorption rate (SAR) is essential for evaluating both hardware and implant safety, yet thermometry and simulation‐only approaches face limitations. In addition, these evaluations are typically performed within the MR scanner, where failures or excessive loading can damage the transmit chain and result in costly repairs and scanner downtime. To overcome these challenges, alternative experimental approaches that enable direct and controlled SAR validation outside the MR environment are needed.

Read PDF

Similar papers

Open access Sep 2026

RF safety assessment of dental retainers for human head MRI array coils at 9.4 T.

A numerical workflow for evaluating radiofrequency (RF) safety for ultra-high field (UHF, 7 T and above) MRI with a local transmit array coil in the presence of conductive objects such as dental retainers is developed.

Egor Berezko, G. Solomakha, N. Avdievich et al. · 0 citations
Open access Sep 2026

Physics-Informed Design and Bench/Phantom Validation of a Shaft-Compatible 13.56 MHz NFC System for Laparoscopic Colorectal Tumour Localisation

Background/Objectives: Accurate intraoperative tumour localisation remains challenging in minimally invasive colorectal surgery because tactile palpation is lost and conventional markers can migrate or provide imprecise localisation. Building on a preceding tri-frequency study that identified 13.56 MHz as the preferred...

Bogdan Mocan, M. Mocan, M. Fulea et al. · 0 citations
Open access Sep 2026

A Fast B 1 + -Driven RF Co-Simulation Optimization Method for RF Coil Arrays.

PURPOSE To generate accurate EM simulation models of RF coil arrays, improve coil safety validation procedures, and improve RF power management without sacrificing patient safety. METHODS A fast, customized RF co-simulation formulation was developed for pTx RF coils, enabling the use of experimentally measured B 1 +...

Mazin M. Mustafa, Ethan Polcyn, Matt Waks et al. · 0 citations
Open access Jan 2026

Quantifying the Transmit‐Power Threshold for Reliable Passive Detuning of Inductively Coupled Wireless Resonators in MRI: A Phantom Study

ABSTRACT This work develops and validates a practical method for quantifying the transmit‐power threshold at which passive detuning of inductively coupled wireless resonators becomes reliable in MRI. All experiments were performed using phantoms. A near/far spatial signal ratio is computed from a series of gradient‐ech...

Ming Lu, Xiao-Yu Jiang, Ya-Jun Ma et al. · 0 citations
Aug 2026

Validation of a Cable-Free and User-Friendly thermoAcoustic Range Verifier for a 2 Gy Proton Field Delivered to the Liver of an Anthropomorphic Phantom.

PURPOSE Detection and offline correction of proton range errors due to incorrect Hounsfield Unit-to-stopping power conversion with minimal (5-10 minute) extension of the patient-facing clinical workflow using a cable-free thermoAcoustic Range Verifier (tARV). MATERIALS The accelerator (Mevion Hyperscan) delivered at...

S. Patch, V. Sarkar, Anjali Chiravuri et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.