RF safety assessment of dental retainers for human head MRI array coils at 9.4 T.

Sep 2026 · Magnetic Resonance Imaging · pp. 110803 · 0 citations · 55 references
Medicine

TL;DR

A numerical workflow for evaluating radiofrequency (RF) safety for ultra-high field (UHF, 7 T and above) MRI with a local transmit array coil in the presence of conductive objects such as dental retainers is developed.

Abstract

In this work we developed a numerical workflow for evaluating radiofrequency (RF) safety for ultra-high field (UHF, 7 T and above) MRI with a local transmit array coil in the presence of conductive objects such as dental retainers. The proposed workflow includes evaluation of the B1+ distribution and peak Specific Absorption Rate (pSAR) for the standard circularly polarized modes, worst-case pSAR modes, and random RF modes. The workflow provides a safety margin, which limits the RF transmit power during MRI scanning. The proposed workflow is applicable to different array coils at different UHF strengths (e.g. 7 T, 9.4 T, 11.7 T, 14 T) as well as to other types of conductive objects (e.g., dental or neural implants). 16-element tight-fit transceiver and larger 16-element transmit-only head arrays were numerically evaluated at 9.4 T in the presence of the dental retainer. Updated safety margins were calculated for both arrays. In addition, we calculated Q-matrix and virtual observation point matrices, which facilitate the real-time SAR-monitoring and can be used for optimization of the B1+ homogeneity and SAR.

Read PDF

Similar papers

Open access Aug 2026

SAR validation of RF coils and implants in a B0‐free RF safety laboratory

Direct SAR measurement provides a reliable and quantitative approach for validating RF coil and implant safety, matching the accuracy of conventional scanner‐based thermometry and simulation methods.

Nur Izzati Huda Zulkarnain, Mazin M. Mustafa, Alireza Sadeghi-Tarakameh et al. · 1 citation
Open access Sep 2026

Head-only MRI of Deep Brain Stimulation devices at 0.5 T: Patient-derived approaches for estimating radiofrequency heating.

The growing prevalence of neurodegenerative diseases has increased the number of patients treated with deep brain stimulation (DBS), many of whom will require magnetic resonance imaging (MRI) during their care. However, radiofrequency (RF)-induced heating of conductive DBS leads during MRI remains a primary safety conc...

A. Louka, Diego F. Martinez, Kat Zaraska et al. · 0 citations
Sep 2026

Diagnostic Performance of a Metasurface Coil Compared With a Commercial 12-Channel Coil for Knee MRI at 1.5 T.

BACKGROUND Knee MRI image quality depends largely on the radiofrequency coil; metasurfaces offer a new approach to improving coil performance. PURPOSE To evaluate the imaging performance and diagnostic capability of a metasurface coil for knee MRI. STUDY TYPE Prospective. PHANTOM/POPULATION A cylindrical phantom,...

Jie Gao, Xin-Xin Li, Zhong-Hai Chi et al. · 1 citation

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.