Sep 2026· Magnetic Resonance Imaging· pp.
110803
· 0 citations· 55 references
Medicine
TL;DR
A numerical workflow for evaluating radiofrequency (RF) safety for ultra-high field (UHF, 7 T and above) MRI with a local transmit array coil in the presence of conductive objects such as dental retainers is developed.
Abstract
In this work we developed a numerical workflow for evaluating radiofrequency (RF) safety for ultra-high field (UHF, 7 T and above) MRI with a local transmit array coil in the presence of conductive objects such as dental retainers. The proposed workflow includes evaluation of the B1+ distribution and peak Specific Absorption Rate (pSAR) for the standard circularly polarized modes, worst-case pSAR modes, and random RF modes. The workflow provides a safety margin, which limits the RF transmit power during MRI scanning. The proposed workflow is applicable to different array coils at different UHF strengths (e.g. 7 T, 9.4 T, 11.7 T, 14 T) as well as to other types of conductive objects (e.g., dental or neural implants). 16-element tight-fit transceiver and larger 16-element transmit-only head arrays were numerically evaluated at 9.4 T in the presence of the dental retainer. Updated safety margins were calculated for both arrays. In addition, we calculated Q-matrix and virtual observation point matrices, which facilitate the real-time SAR-monitoring and can be used for optimization of the B1+ homogeneity and SAR.
Direct SAR measurement provides a reliable and quantitative approach for validating RF coil and implant safety, matching the accuracy of conventional scanner‐based thermometry and simulation methods.
Nur Izzati Huda Zulkarnain, Mazin M. Mustafa, Alireza Sadeghi-Tarakameh et al.· Medical Physics (Lancaster)· 1 citation
The growing prevalence of neurodegenerative diseases has increased the number of patients treated with deep brain stimulation (DBS), many of whom will require magnetic resonance imaging (MRI) during their care. However, radiofrequency (RF)-induced heating of conductive DBS leads during MRI remains a primary safety conc...
A. Louka, Diego F. Martinez, Kat Zaraska et al.· NeuroImage· 0 citations
The flexible blanket-like torso coil provided image quality similar to a conventional coil and was generally preferred by patients and MRI technologists.
J. Dillman, Jean A. Tkach, R. Guillerman et al.· Pediatric Radiology· 0 citations
BACKGROUND
Knee MRI image quality depends largely on the radiofrequency coil; metasurfaces offer a new approach to improving coil performance.
PURPOSE
To evaluate the imaging performance and diagnostic capability of a metasurface coil for knee MRI.
STUDY TYPE
Prospective.
PHANTOM/POPULATION
A cylindrical phantom,...
Jie Gao, Xin-Xin Li, Zhong-Hai Chi et al.· Journal of Magnetic Resonanc...· 1 citation
This work presents an open-source, optimized solenoid head coil tailored for the 50 mT open-source scanner (OSII ONE v2.1), set the basis for a fully reliable and reproducible component for the open-source OSII ONE MRI scanner.
Umberto Zanovello, Julia Pfitzer, Ariane Ernst et al.· 1 citation