Aug 2026· European Journal of Radiology· Vol 205, pp.
113177
· 0 citations· 24 references
Medicine
Abstract
Purpose
To evaluate an iterative motion-correction algorithm for periinterventional hepatic cone-beam computed tomography (CBCT) regarding its efficacy in reducing motion artifacts, improving vessel depiction, and enhancing diagnostic confidence during transarterial interventions.
Materials And Methods
This retrospective single-centre study included 69 CBCT datasets from 69 patients undergoing TACE or SIRT between 2018 and 2021. One CBCT dataset per patient served as the unit of analysis. Each dataset was post-processed with a motion-correction algorithm at 100-iteration increments (It0 = Baseline; It100-It1000). Three interventional radiologists independently assessed motion artifacts (MA) and vessel depiction (VD) using a 5-point Likert scale (higher ratings representing better image quality). Newly generated artifacts (NGA) were assessed on a 0-5 scale ("0″ representing no evidence of NGA, higher ratings representing stronger artifacts). Target lesion (TL) and vessel tree (VT) visibility were defined as yes/no. Non-parametric tests were applied, including Friedman and Wilcoxon signed-rank tests for ordinal ratings and Cochran's Q and McNemar tests for binary visibility outcomes. Interrater reliability was determined using Fleiss' kappa.
Results
Compared with baseline datasets, median ratings for motion artifacts and vessel depiction improved from 3 (IQR 1-2) to 4 (IQR 1) at iteration levels 200-300 (all p ≤ 0.001). Target lesion and vessel tree visibility increased from baseline values of 42-44% and 30-41% to peak rates of 65-75% and 54-64%, respectively, with significant differences across iteration levels for all readers (all p < 0.001). Beyond 500 iterations, diagnostic quality declined due to progressively increasing NGA. The proportion of non-diagnostic datasets increased continuously with higher iteration levels, ranging from 1.4% at It300 to 88.4% at It1000. Interrater agreement across parameters was moderate to substantial (κ = 0.40-0.70).
Conclusion
Iterative motion correction significantly improves image quality in hepatic CBCT during transarterial interventions at moderate iteration levels (200-300). Excessive iteration introduces new artifacts, underscoring the importance of optimizing mid-range iteration settings to balance motion correction and artifact generation.
BACKGROUND
Cone-beam computed tomography (CBCT) is a critical component of liver directed therapies used to identify tumor feeding vessels. CBCT images are often limited by breathing motion artifacts. This study evaluates if post-processing CBCT images with a motion compensation algorithm improves image-quality for tra...
Katherine L. McNamara, Austin Ng, A. Vezeridis et al.· MITAT. Minimally invasive th...· 0 citations
MPR image quality is typically inferior to axial images and deteriorates further when derived from thicker axial slices, therefore, appropriate selection of the reconstruction kernel, pixel size and slice thickness are essential to maintain diagnostic image quality.
P. Monnin, A. Viry, F. Becce et al.· Journal of Applied Clinical...· 0 citations
The findings support implementing standardized protocols with automatic dual-range reconstructions to enable efficient, individualized carotid plaque assessment in clinical practice and support implementing standardized protocols with automatic dual-range reconstructions for keV selection based on plaque composition.
Mirela Kostadinova, Lea B. Uebelacker, Tommaso D'Angelo et al.· Diagnostics· 0 citations
Subject motion during MRI acquisition is one of the major factors limiting image quality and diagnostic value. Previous studies have demonstrated the effectiveness of prospective motion correction (PMC) in mitigating motion‐related artifacts. However, the effectiveness of PMC for the visualization of perivascular space...
Heng-Heng Wu, Bing-Bing Zhao, Cao-Hui Duan et al.· NMR in Biomedicine· 0 citations
Background: Noise reduction in low-dose pediatric computed tomography (CT) requires effective suppression of quantum noise without compromising subtle anatomical structures. This study compared three bilateral-based noise reduction algorithms, namely the conventional bilateral filter (BF), adaptive BF (ABF), and traina...
Minji Park, Hajin Kim, Sang Woong Park et al.· Diagnostics· 0 citations
The motorized attachment for SR3D acquisition reduced image variability and improved image quality, and may encourage broader use of sagittal reconstruction in HDR-BT, potentially improving patient outcomes.
David Contella, Chandima Edirisinghe, Kevin Barker et al.· Brachytherapy· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.