Comparability of serum folate concentration measured by a microbiological method, LC-MS/MS and immunoassay.

Sep 2026 · Annals of Clinical Biochemistry · pp. 45632261488256 · 0 citations
Medicine

TL;DR

Data shows that the microbiological method, LC-MS/MS and Tosoh CL1200 immunoassay have acceptable accuracy, imprecision and agreement, and the small concentration dependant bias is likely explainable by the expected characteristics of the individual methods and ability to detect and respond to different forms of folate.

Abstract

Background Folate, vitamin B9, is an essential micronutrient for DNA replication and repair. Deficiency is associated with megaloblastic anaemia and fetal neural tube defects. Serum folate measurement provides a convenient biomarker of folate status in clinical and research settings. The overall aim of this evaluation was to compare serum folate results measured from three different technologies within the same laboratory to determine systematic biases and equivalence between methods. Methods 58 serum samples from the UK National Diet and Nutrition Survey were used to compare a microbiological method, a liquid chromatography tandem mass spectrometry method (LC-MS/MS) and a commercial immunoassay (Tosoh CL1200). Results Within-run imprecision of all three assays was ≤7%, measured at two concentrations. The microbiological and LC-MS/MS methods were accurate against NIST Standard Reference Material and VITAL External Quality Assessment (EQA) target values. The plasma-based sample used for the NIST Standard Reference Material (SRM) was unsuitable for analysis by the Tosoh immunoassay, however comparison with the VITAL EQA target values showed good accuracy. Using the serum samples, both the LC-MS/MS and Tosoh immunoassay methods showed a slight positive proportional bias compared to the microbiological method, whilst the Tosoh method also had an additional constant bias of 3-4 nmol/L compared to the other methods. Conclusions This data shows that the microbiological method, LC-MS/MS and Tosoh CL1200 immunoassay have acceptable accuracy, imprecision and agreement. The small concentration dependant bias is likely explainable by the expected characteristics of the individual methods and ability to detect and respond to different forms of folate.

Read PDF

Similar papers

Oct 2026

B-370 Toward the Standardization of Homocysteine Testing: A Metrological Framework Based on a Candidate Reference Method and Commutable Candidate Reference Materials

Homocysteine (Hcy) is a critical biomarker for cardiovascular risk stratification. However, significant variability exists among routine detection methods in clinical laboratories across China, as evidenced by external quality assessment data, underscoring an urgent need for standardization to ensure reliable and...

Wei-Yan Zhou · 0 citations
Oct 2026

B-415 Agreement Between Colorimetric Roche Cobas PRO and ICP-MS Zinc Assays in Blood Samples

Accurate measurement of zinc concentrations in blood is essential for assessing nutritional status and diagnosing deficiency. This study compared zinc results obtained using the Roche Cobas Pro colorimetric assay with those generated by inductively coupled plasma mass spectrometry (ICP-MS), the analytical referen...

Waleed Tamimi, Khaled Ahmad, M. Elsafdi et al. · 0 citations
Open access Sep 2026

Comprehensive Comparison of Three Automated 25-Hydroxyvitamin D Immunoassays with a VDSCP Traceable LC-MS/MS Method Across Clinically Relevant Decision Thresholds

Accurate measurement of serum 25-hydroxyvitamin D (25OHD) is essential for the diagnosis and management of vitamin D deficiency and the importance of method-aware interpretation of vitamin D results is highlighted to support continued assay harmonization to improve comparability in routine clinical practice.

Szilvia Racz, Luca Jozsa, Amrit Pal Bhattoa-Buzas et al. · 0 citations
Open access Aug 2026

Preliminary analytical evaluation of the fully automated Cobas i601 LC-MS/MS system for steroid hormones and vitamin D in a clinical laboratory.

BACKGROUND Fully automated liquid chromatography-tandem mass spectrometry (LC-MS/MS) platforms aim to combine the analytical specificity and sensitivity of mass spectrometry with the workflow efficiency of automated routine analyzers. This study performed a preliminary analytical evaluation of the Cobas i601 LC-MS/MS s...

Mariana Serres-Gómez, María Luisa González-Casaus, María Gemma Serrano Olmedo et al. · 0 citations
Open access Oct 2026

Analytical Performance of Vitamin D Immunoassays Compared with Electrochemiluminescence Immunoassay: A Cross-sectional Study

Introduction: Vitamin D, the fat-soluble steroid hormone, is important in maintaining homeostasis of calcium and phosphorus. The diagnosis of Vitamin D deficiency has become considerably more frequent due to its importance in skeletal and non skeletal disorders. Comparison among immunoassay methods is essential as meth...

Arun Radhakrishnan, Ajaikumar Sukumaran, Haritha Raveendran et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.