Sep 2026· Annals of Clinical Biochemistry· pp.
45632261488256
· 0 citations
Medicine
TL;DR
Data shows that the microbiological method, LC-MS/MS and Tosoh CL1200 immunoassay have acceptable accuracy, imprecision and agreement, and the small concentration dependant bias is likely explainable by the expected characteristics of the individual methods and ability to detect and respond to different forms of folate.
Abstract
Background Folate, vitamin B9, is an essential micronutrient for DNA replication and repair. Deficiency is associated with megaloblastic anaemia and fetal neural tube defects. Serum folate measurement provides a convenient biomarker of folate status in clinical and research settings. The overall aim of this evaluation was to compare serum folate results measured from three different technologies within the same laboratory to determine systematic biases and equivalence between methods. Methods 58 serum samples from the UK National Diet and Nutrition Survey were used to compare a microbiological method, a liquid chromatography tandem mass spectrometry method (LC-MS/MS) and a commercial immunoassay (Tosoh CL1200). Results Within-run imprecision of all three assays was ≤7%, measured at two concentrations. The microbiological and LC-MS/MS methods were accurate against NIST Standard Reference Material and VITAL External Quality Assessment (EQA) target values. The plasma-based sample used for the NIST Standard Reference Material (SRM) was unsuitable for analysis by the Tosoh immunoassay, however comparison with the VITAL EQA target values showed good accuracy. Using the serum samples, both the LC-MS/MS and Tosoh immunoassay methods showed a slight positive proportional bias compared to the microbiological method, whilst the Tosoh method also had an additional constant bias of 3-4 nmol/L compared to the other methods. Conclusions This data shows that the microbiological method, LC-MS/MS and Tosoh CL1200 immunoassay have acceptable accuracy, imprecision and agreement. The small concentration dependant bias is likely explainable by the expected characteristics of the individual methods and ability to detect and respond to different forms of folate.
Homocysteine (Hcy) is a critical biomarker for cardiovascular risk stratification. However, significant variability exists among routine detection methods in clinical laboratories across China, as evidenced by external quality assessment data, underscoring an urgent need for standardization to ensure reliable and...
Accurate measurement of zinc concentrations in blood is essential for assessing nutritional status and diagnosing deficiency. This study compared zinc results obtained using the Roche Cobas Pro colorimetric assay with those generated by inductively coupled plasma mass spectrometry (ICP-MS), the analytical referen...
Waleed Tamimi, Khaled Ahmad, M. Elsafdi et al.· Clinical Chemistry· 0 citations
Accurate measurement of serum 25-hydroxyvitamin D (25OHD) is essential for the diagnosis and management of vitamin D deficiency and the importance of method-aware interpretation of vitamin D results is highlighted to support continued assay harmonization to improve comparability in routine clinical practice.
Szilvia Racz, Luca Jozsa, Amrit Pal Bhattoa-Buzas et al.· Nutrients· 0 citations
BACKGROUND
Fully automated liquid chromatography-tandem mass spectrometry (LC-MS/MS) platforms aim to combine the analytical specificity and sensitivity of mass spectrometry with the workflow efficiency of automated routine analyzers. This study performed a preliminary analytical evaluation of the Cobas i601 LC-MS/MS s...
Mariana Serres-Gómez, María Luisa González-Casaus, María Gemma Serrano Olmedo et al.· Clinica chimica acta; intern...· 0 citations
Introduction: Vitamin D, the fat-soluble steroid hormone, is important in maintaining homeostasis of calcium and phosphorus. The diagnosis of Vitamin D deficiency has become considerably more frequent due to its importance in skeletal and non skeletal disorders. Comparison among immunoassay methods is essential as meth...
Arun Radhakrishnan, Ajaikumar Sukumaran, Haritha Raveendran et al.· Journal of Clinical and Diag...· 0 citations
DiaSorin LIASON XL 25-OH Total vitamin D immunoassay demonstrated acceptable and good analytical performance during assay verification and after implementation as part of the monitoring of patient testing.
J. Jara-Aguirre, Charity Tabuno, Charles D. Cottone et al.· Clinical Chemistry· 0 citations