Comprehensive Comparison of Three Automated 25-Hydroxyvitamin D Immunoassays with a VDSCP Traceable LC-MS/MS Method Across Clinically Relevant Decision Thresholds

Sep 2026 · Nutrients · Vol 18, pp. 2986 · 0 citations · 44 references
Medicine

TL;DR

Accurate measurement of serum 25-hydroxyvitamin D (25OHD) is essential for the diagnosis and management of vitamin D deficiency and the importance of method-aware interpretation of vitamin D results is highlighted to support continued assay harmonization to improve comparability in routine clinical practice.

Abstract

Background/Objective: Accurate measurement of serum 25-hydroxyvitamin D (25OHD) is essential for the diagnosis and management of vitamin D deficiency. Although automated immunoassays are widely used in routine clinical practice, analytical variability between platforms may lead to inconsistent patient classification around clinically relevant decision thresholds. This study comprehensively compared three automated immunoassays (DiaSorin Liaison XL, Beckman Coulter DxI 9000, and Roche cobas e602) with a Vitamin D Standardization Certification Program (VDSCP)-traceable liquid chromatography–tandem mass spectrometry (LC-MS/MS) reference method. Methods: A total of 400 anonymized adult outpatient serum samples were prospectively collected and equally distributed across four predefined vitamin D concentration categories (<25, 25–49, 50–74, and ≥75 nmol/L). Samples were stored at −70 °C before simultaneous analysis on the three automated immunoassays and the reference LC-MS/MS platform. Method comparison included Deming regression, Bland–Altman analysis, and evaluation of diagnostic performance at clinically relevant thresholds (<25, <50, and <75 nmol/L). The effects of sample storage, vitamin D metabolites (C3-epi-25OHD3 and 24,25-dihydroxyvitamin D3), and external quality assessment (DEQAS) performance were also investigated. Results: Significant analytical differences were observed between all immunoassays and LC-MS/MS. Roche cobas e602 consistently overestimated 25OHD concentrations, resulting in reduced sensitivity but excellent specificity and positive predictive value across all clinical thresholds. Beckman Coulter demonstrated the highest sensitivity for detecting vitamin D deficiency but generated more false-positive classifications, particularly at the <75 nmol/L threshold. DiaSorin Liaison XL achieved the best overall balance between sensitivity, specificity, accuracy, and agreement with LC-MS/MS, with Cohen’s κ values ranging from 0.818 to 0.895 across decision thresholds. Higher concentrations of C3-epi-25OHD3 and 24,25-dihydroxyvitamin D3 were associated with greater positive bias in Roche measurements, while Beckman Coulter showed no significant metabolite-related interference, although the present correlation analyses could not establish independent analytical interference. DEQAS performance did not consistently reflect agreement with LC-MS/MS in patient samples. Conclusions: Despite ongoing international standardization efforts, clinically important differences remain among automated 25OHD immunoassays. Assay-specific analytical characteristics significantly influence patient classification around established decision thresholds. DiaSorin Liaison XL demonstrated the closest overall agreement with the reference LC-MS/MS method, whereas Beckman Coulter prioritized sensitivity and Roche specificity. These findings highlight the importance of method-aware interpretation of vitamin D results and support continued assay harmonization to improve comparability in routine clinical practice.

Read PDF

Similar papers

Open access Aug 2026

Preliminary analytical evaluation of the fully automated Cobas i601 LC-MS/MS system for steroid hormones and vitamin D in a clinical laboratory.

BACKGROUND Fully automated liquid chromatography-tandem mass spectrometry (LC-MS/MS) platforms aim to combine the analytical specificity and sensitivity of mass spectrometry with the workflow efficiency of automated routine analyzers. This study performed a preliminary analytical evaluation of the Cobas i601 LC-MS/MS s...

Mariana Serres-Gómez, María Luisa González-Casaus, María Gemma Serrano Olmedo et al. · 0 citations
Sep 2026

Analytical performance of a single epitope meta-sandwich total 25-hydroxyvitamin D immunoassay and establishment of reference intervals in a Chinese population.

BACKGROUND 25-hydroxyvitamin D [25-(OH)D] concentration is considered to evaluate vitamin D status. Persistent discrepancies between chemiluminescent immunoassays (CLIAs) and liquid chromatography-tandem mass spectrometry (LC-MS/MS) continue to challenge clinical laboratories. To investigate the single epitope meta-san...

Xiao-Chun Dai, Chun-Man Xia, Fang Zhang et al. · 0 citations
Oct 2026

B-370 Toward the Standardization of Homocysteine Testing: A Metrological Framework Based on a Candidate Reference Method and Commutable Candidate Reference Materials

Homocysteine (Hcy) is a critical biomarker for cardiovascular risk stratification. However, significant variability exists among routine detection methods in clinical laboratories across China, as evidenced by external quality assessment data, underscoring an urgent need for standardization to ensure reliable and...

Wei-Yan Zhou · 0 citations
Oct 2026

B-365 Breaking Throughput Barriers: Automated Vitamin D Testing on the Roche Cobas i601 LC-MS/MS in a Tertiary Hospital Laboratory

Measurement of 25-hydroxyvitamin D (25-OH-D), including 25-OH-D2 and 25-OH-D3, is essential for assessing vitamin D status in clinical practice. Liquid chromatography–tandem mass spectrometry (LC-MS/MS) is considered the reference method due to its superior analytical specificity and ability to distinguish vitami...

S. Wong, C. Yeo, C. Gea et al. · 0 citations
Oct 2026

Analytical performance and clinical classification value of a novel sandwich immunoassay for total testosterone measurement in low-concentration and hemodialysis populations

Abstract Objectives To evaluate the analytical performance and clinical classification value of a novel sandwich testosterone immunoassay, particularly for biochemical testosterone elevation in women and deficiency in male hemodialysis patients, using LC–MS/MS as the reference method. Methods A total of 387 residual cl...

He Shi, Hui-Mei Wei, Shuai-Wei Han et al. · 0 citations
Open access Oct 2026

Analytical Performance of Vitamin D Immunoassays Compared with Electrochemiluminescence Immunoassay: A Cross-sectional Study

Introduction: Vitamin D, the fat-soluble steroid hormone, is important in maintaining homeostasis of calcium and phosphorus. The diagnosis of Vitamin D deficiency has become considerably more frequent due to its importance in skeletal and non skeletal disorders. Comparison among immunoassay methods is essential as meth...

Arun Radhakrishnan, Ajaikumar Sukumaran, Haritha Raveendran et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.