The dual-role coil demonstrated a high degree of flexibility in controlling the transmit field and reducing RF-induced heating at DBS implants, offering a novel approach to mitigate RF-induced heating of the implants in MRI.
Abstract
Magnetic resonance imaging (MRI) scanning remains largely restricted to specific modalities, typically involving low radiofrequency (RF) power levels and stringent protocols for patients with deep brain stimulation (DBS) implants, due to safety concerns related to RF-induced heating of the implants. A 6-channel dual-role head coil array capable of modulating the electric-field (E-field) distribution was designed and evaluated using electromagnetic (EM) simulations. By optimizing the resonant frequency of each coil element during RF transmission, the transmit field was reshaped, leading to a significant reduction in RF-induced heating near the DBS lead tip. The proposed method was validated across two scenarios of increasing complexity: 1) a simple straight conductive wire for concept validation and 2) four realistic DBS leads representing complex real-world scenarios. The coil settings can be optimized either to suppress the E-field at a specific location, such as the DBS lead tip, or to suppress the peak specific absorption rate (SAR) across the entire human head. For location-specific E-field suppression, the simplified predefined-state control scheme and the fine-tuning genetic algorithm (GA)-based framework were implemented, achieving E-field reductions of 45.9% and 68.3%, respectively. For whole-head peak SAR suppression, the annealed Log-Sum-Exp (LSE)–Adaptive Moment Estimation (Adam) framework (LSE–Adam) was implemented, achieving an average 1 g SAR reduction of 72.09% across four realistic DBS lead models. The dual-role coil demonstrated a high degree of flexibility in controlling the transmit field and reducing RF-induced heating at DBS implants, offering a novel approach to mitigate RF-induced heating of the implants in MRI.
Axial conductivity tapering markedly attenuates MRI-induced RF heating without requiring significant changes to lead geometry and provide a practical path toward inherently MRI-compatible AIMDs, potentially expanding diagnostic imaging access for millions of patients.
Tayeb Zaidi, F. Marturano, P. Sanpitak et al.· IEEE transactions on bio-med...· 3 citations· ⚡1
This work presents an open-source, optimized solenoid head coil tailored for the 50 mT open-source scanner (OSII ONE v2.1), set the basis for a fully reliable and reproducible component for the open-source OSII ONE MRI scanner.
Umberto Zanovello, Julia Pfitzer, Ariane Ernst et al.· 1 citation
This review elaborates on the inductive coupling mechanism of ICWCs, the derivation of the SNR formula, potential causes of g-factor reduction, and recommendations for fabrication methods, providing a reference for the innovation of MRI RF coil technology and the clinical translation of ICWCs.
High-performance breast magnetic resonance imaging (MRI) remains limited by the availability, geometry, and cost of dedicated receive coils. In contrast, spine coils are widely installed in MR scanners globally but provide insufficient sensitivity for anterior breast tissue. Here, we report a wireless volumetric metama...
Yuhang Liu, Xia Zhu, C. LeBedis et al.· 0 citations
Direct SAR measurement provides a reliable and quantitative approach for validating RF coil and implant safety, matching the accuracy of conventional scanner‐based thermometry and simulation methods.
Nur Izzati Huda Zulkarnain, Mazin M. Mustafa, Alireza Sadeghi-Tarakameh et al.· Medical Physics (Lancaster)· 1 citation