A Novel Lead Construct to Reduce MRI-Induced RF Heating: Construction, In Vitro Validation, and In Vivo Predictions.

Jul 2026 · IEEE transactions on bio-medical engineering · Vol PP, pp. 1-12 · 3 citations · ⚡ 1 influential
Medicine

TL;DR

Axial conductivity tapering markedly attenuates MRI-induced RF heating without requiring significant changes to lead geometry and provide a practical path toward inherently MRI-compatible AIMDs, potentially expanding diagnostic imaging access for millions of patients.

Abstract

Objective

Magnetic resonance imaging (MRI) of patients with active implantable medical devices is restricted by radiofrequency (RF) heating of conductive leads. Our goal was to develop and validate resistively tapered cylindrical (RTC) leads that intrinsically suppress RF heating.

Methods

Two-segment RTC wire prototypes (RTC1, RTC2) were fabricated by thin-film physical vapor deposition to create axial conductivity discontinuities. The transfer functions for both were measured and validated with in vitro heating experiments at 1.5 T along 12 trajectories and benchmarked against a uniform conductivity control wire. Calibrated transfer functions were combined with electromagnetic simulations of 210 clinically realistic deep brain stimulation (DBS) trajectories in an anatomically detailed full-body model to predict in vivo heating, with a commercial DBS lead included for comparison.

Results

Gel phantom experiments showed the RTC wires cut peak RF-induced temperature rise at the tip by $>$60% versus the control wire in vitro, lowering the mean heating from 3.67 to 1.43. The in vivo predictions using the transfer function methodology showed a $>$20% reduction in the mean tip heating vs. the uniform conductivity wire ($p< 0.001$) and a 20-fold lower peak heating compared to a commercial DBS lead under identical exposure conditions.

Conclusion

Axial conductivity tapering markedly attenuates MRI-induced RF heating without requiring significant changes to lead geometry.

Significance

Conductivity-tailored leads provide a practical path toward inherently MRI-compatible AIMDs, potentially expanding diagnostic imaging access for millions of patients.

Read PDF

Similar papers

Open access Sep 2026

Head-only MRI of Deep Brain Stimulation devices at 0.5 T: Patient-derived approaches for estimating radiofrequency heating.

The growing prevalence of neurodegenerative diseases has increased the number of patients treated with deep brain stimulation (DBS), many of whom will require magnetic resonance imaging (MRI) during their care. However, radiofrequency (RF)-induced heating of conductive DBS leads during MRI remains a primary safety conc...

A. Louka, Diego F. Martinez, Kat Zaraska et al. · 0 citations
Preprint Oct 2026

A Guidewire Trajectory Generation Tool for Interventional MRI Simulations

Significance: RF-induced heating of endovascular devices limits MRI-guided interventions, including targeted drug delivery to brain tumors. Purpose: To present a tool that rapidly generates randomized guidewire trajectories for RF heating risk assessment. Methods: A MATLAB tool models the guidewire as a spline with ran...

Fatima Ben Haj, Jeffrey W. Hand, S. Malik et al. · 0 citations
Open access Aug 2026

SAR validation of RF coils and implants in a B0‐free RF safety laboratory

Direct SAR measurement provides a reliable and quantitative approach for validating RF coil and implant safety, matching the accuracy of conventional scanner‐based thermometry and simulation methods.

Nur Izzati Huda Zulkarnain, Mazin M. Mustafa, Alireza Sadeghi-Tarakameh et al. · 1 citation
Open access Sep 2026

RF safety assessment of dental retainers for human head MRI array coils at 9.4 T.

A numerical workflow for evaluating radiofrequency (RF) safety for ultra-high field (UHF, 7 T and above) MRI with a local transmit array coil in the presence of conductive objects such as dental retainers is developed.

Egor Berezko, G. Solomakha, N. Avdievich et al. · 0 citations
Preprint Sep 2026

A wearable, stretchable radio-frequency coil for extremity imaging in low-field MRI systems

Purpose: To develop a wearable, stretchable radio-frequency (RF) coil that conforms to the extremities and improves the filling factor in low-field MRI. Methods: A stretchable solenoid RF coil was fabricated by stitching a sinusoidally arranged Litz-wire conductor onto an elastic textile. The coil was compared with fou...

J. Conejero, Marina Fernández-García, J. Borreguero et al. · 0 citations
Open access Aug 2026

Design, verification and validation of a spatially confined magnetic nanoparticle thermal therapy device for conformal thermal dose contouring

Background: Magnetic nanoparticle thermal therapy (MNTT) can localize heat within tumors, but non-uniform nanoparticle distribution can limit thermal dose coverage. Spatially confined MNTT (SC-MNTT) addresses this limitation by using a static magnetic field (SMF) to suppress heating outside a selected field-free region...

Y. Lad, Shreeniket Pawar, T. Townsend et al. · 0 citations