A Specialized Lipids Panel LDT on the Roche Cobas 8000 c702 that integrates precipitation-based HDL-C with ultracentrifugation-derived LDL-C provides a more accurate alternative to estimation formulas and direct homogenous assays, particularly in complex lipid profiles.
Abstract
HDL-C and LDL-C are key cardiovascular biomarkers, but traditional estimation formulas lose accuracy in high triglyceride and non-fasting states, and direct homogeneous assays remain vulnerable to matrix interferences. Although ultracentrifugation is considered the reference method for LDL-C measurement, its technical complexity and labor-intensive workflow make it impractical for routine clinical use. These limitations are particularly consequential in clinical trials, where precise lipid quantification is critical for evaluating therapeutic efficacy. To address these challenges, Labcorp developed a Specialized Lipids Panel LDT on the Roche Cobas 8000 c702 that integrates precipitation-based HDL-C with ultracentrifugation-derived LDL-C, providing a higher accuracy alternative to conventional lipid measurements.
This Laboratory Developed Test (LDT) panel measures HDL -C isolated from neat serum using dextran sulfate precipitation and quantified with the Roche Total cholesterol assay. Total cholesterol in the ultracentrifugation bottom fraction was also quantified using the Roche assay. LDL-C was calculated by subtracting HDL-C from the cholesterol concentration of the bottom fraction. Method validation followed Labcorp SOPs and CLSI guidelines, assessing linearity, precision, accuracy, sensitivity, specificity, dilution linearity, and stability. EP Evaluator® was used for statistical analysis.
Analytical measuring ranges were 0.260–5.170 mmol/L (HDL-C) and 0.260–20.690 mmol/L (LDL-C). Intra assay CVs were =4.1% (HDL-C) and =5.2% (LDL-C); inter assay CVs =5.4% and =6.8%. Accuracy was demonstrated with acceptable CDC reference and spike and recovery results. The LLOQ for both analytes was 0.260 mmol/L. No clinically significant interference was observed from hemoglobin (=862 mg/dL HDL-C; =711 mg/dL LDL-C), triglycerides (=0.3 mg/dL; =46 mg/dL), conjugated bilirubin (=20 mg/dL; =7 mg/dL), or unconjugated bilirubin (=42 mg/dL; =10 mg/dL). Manual dilution to 1:16 extended ULOQs to 82.720 mmol/L (HDL-C) and 331.040 mmol/L (LDL-C). Stability was confirmed for 14 days refrigerated and at -70?°C.
The Specialized Lipids Panel LDT demonstrated robust analytical performance and meets Labcorp’s acceptance criteria for clinical use. By leveraging precipitation-based HDL-C and ultracentrifugation-derived LDL-C, this LDT provides a more accurate alternative to estimation formulas and direct homogenous assays, particularly in complex lipid profiles.
Cardiovascular disease continues to be the universal leading cause of death. Preventative screening and monitoring plays an important role in mitigating risk, however, socially disadvantaged communities frequently face major challenges in receiving appropriate care. The evolution of cardiovascular risk assessment...
Anissa R. Naranjo, Carlos Torres, C. Carter et al.· Clinical Chemistry· 0 citations
Low-density lipoprotein cholesterol (LDL-C) remains the primary target for cardiovascular risk reduction; however, it reflects cholesterol content rather than the total number of circulating atherogenic particles. Apolipoprotein B (ApoB), present in one copy per atherogenic lipoprotein particle, more directly rep...
M. Bezerra, Alice Petrola Bezerra Sayao, Rachel Petrola et al.· Clinical Chemistry· 0 citations
AIM
Lipoprotein(a) [Lp(a)] is an independent risk factor for cardiovascular disease (CVD); however, the goal of resolving its measurement issues remains unachieved. Expanding on the "first step" by Miida et al., this study evaluated the feasibility of harmonizing Lp(a) measurements using an IFCC endorsed mass spectrome...
Keita Yamashita, Mitsuru Wakita, Noriyasu Niizeki et al.· Journal of atherosclerosis a...· 1 citation
Abstract Objectives Accurate estimation of low-density lipoprotein cholesterol (LDL-C) is essential for cardiovascular risk management. This study evaluated the performance of the Friedewald, Martin–Hopkins (MH), and National Institutes of Health Equation 2 (NIH2) methods compared with direct enzymatic measurement acro...
M. Ağıllı, Alpaslan Coşar, E. Karaismailoğlu et al.· Turkish Journal of Biochemis...· 0 citations
Comprehensive lipoprotein analysis and interpretation is a cornerstone of cardiovascular disease (CVD) prevention. Often, multiple platforms and assays are used to quantify lipoprotein parameters and apolipoproteins. Growing economic pressure in healthcare systems is driving the need for efficient technologies ca...
Julia Kurak, J. Meeusen, L. Donato et al.· Clinical Chemistry· 0 citations
The established multi-dimensional lipid dissociation technology effectively degrades lipid particles, mitigates chylomicron interference, and eliminates matrix effects in lipemic samples, overcoming the long-standing challenge of lipemia interference.