Lead-to-Vein Ratio and Long-Term Vascular Safety After Transvenous Pacing in Children Weighing ≤15 Kg.

Sep 2026 · Pacing and clinical electrophysiology : PACE · 0 citations · 24 references
Medicine

TL;DR

During the first transvenous system lifespan, imaging-defined venous abnormalities were relatively frequent but mostly asymptomatic, whereas clinically overt vascular complications were uncommon, suggesting the lead-to-vein ratio may support anatomical pre-implant risk stratification.

Abstract

Background

Transvenous pacing in low-weight children is technically feasible, but its clinical use depends on long-term venous patency, system performance, and ventricular function. We aimed to describe imaging-defined venous abnormalities and clinically overt vascular events after first transvenous pacemaker implantation and to explore whether the lead-to-vein ratio provides anatomical information beyond body weight.

Methods

This prospective observational cohort included 35 children weighing ≤15 kg undergoing first transvenous pacemaker implantation between 2010 and 2020. The primary vascular endpoint was imaging-defined venous thrombosis, occlusion, or stenosis ≥50%; clinical impact was recorded separately. Device-related safety, minimum venous diameter, lead-to-vein ratio, ventricular function, synchrony, and paced QRS duration were analyzed using non-parametric, exact, and exploratory univariable tests.

Results

Median follow-up was 6.7 [4.5-8.2] years. Eleven patients weighed <10 kg and 3 weighed <5 kg. The imaging-defined vascular endpoint occurred in 6/35 patients; 5 were asymptomatic surveillance findings and 1 was clinically symptomatic. The lead-to-vein ratio was higher in patients with the endpoint than in those without events (0.59 [0.55-0.63] vs. 0.47 [0.43-0.51]; p = 0.003). A ratio >0.55 was associated with more frequent imaging-defined endpoints (4/8 vs. 2/27; odds ratio 12.5; 95% confidence interval 1.7-92.3). Left ventricular ejection fraction remained preserved.

Conclusion

During the first transvenous system lifespan, imaging-defined venous abnormalities were relatively frequent but mostly asymptomatic, whereas clinically overt vascular complications were uncommon. The lead-to-vein ratio may support anatomical pre-implant risk stratification, although multicenter validation and follow-up beyond generator replacement are required.

View source

Similar papers

Open access Sep 2026

Long-term cardiac remodeling after transjugular intrahepatic portosystemic shunt placement: a cardiac MRI study.

OBJECTIVES To evaluate long-term cardiovascular effects following insertion of a transjugular intrahepatic portosystemic shunt (TIPS) in patients with cirrhosis-induced portal hypertension. MATERIALS AND METHODS In this prospective intra-individual study from 11/2018 to 5/2022, patients underwent combined hepatic and...

Thomas M. Vollbrecht, N. Mesropyan, Johannes Chang et al. · 0 citations
Open access Oct 2026

Evaluation of Intraprocedural Strategies to Assess Acute Pulmonary Vein Isolation After First-Pass Wide Antral Circumferential Ablation

Background: Atrial fibrillation (AF) is the most common sustained cardiac arrhythmia. While pulmonary vein isolation (PVI) remains the cornerstone of AF ablation, the optimal method for confirming complete acute isolation remains uncertain. This study evaluated the diagnostic value of pulmonary vein (PV) pacing compare...

L. Katov, Julian Ballweg, Y. Teumer et al. · 0 citations
Open access Oct 2026

Long-term changes in pacemaker stimulation burden and its influence on clinical outcomes after transcatheter aortic valve implantation: a multicenter study

It is indicated that patients with persistently high pacing burden should may benefit from structured cardiological follow-up, including optimization of pacing parameter, and a decreasing pacing rate (high/low) was significantly associated with a lower risk of MACE.

R. Schmitt, F. Rees, P. Steinhoff et al. · 0 citations
Open access Sep 2026

Leadless Pacemaker Implant Location as a Predictor of Pacing Induced Cardiomyopathy.

BACKGROUND Leadless pacemakers (LPs) avoid the lead- and pocket-related complications of transvenous systems, but they pace exclusively from the right ventricle (RV), and chronic RV pacing carries a recognized risk of pacing-induced cardiomyopathy (PICM). Whether implant location within the RV, a potentially modifiable...

A. Kushnir, Mones M. Aboelatta, R. Donnino et al. · 0 citations
Nov 2026

Efficacy and Safety of Routine Postdilation in Balloon-Expandable Transcatheter Aortic Valve Implantation Platform: The DUOTAP Randomized Clinical Trial.

Importance In patients with severe aortic stenosis (AS) undergoing transcatheter aortic valve implantation (TAVI), asymmetric valve expansion can estimate prosthetic dysfunction and death. Postdilation has shown promising results in optimizing valve geometry in observational studies; however, randomized evidence is cur...

M. Autherith, K. Singer, L. Hauptmann et al. · 1 citation

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.