Comparison of Blood Glucose Measurement Methods at Low Concentrations in Neonates at Risk for Hypoglycemia.

Sep 2026 · The Journal of Applied Laboratory Medicine · 0 citations
Medicine

TL;DR

At low glucose concentrations, agreement between the Accu-Chek and ABL90 was variable and bidirectional, while agreement between the BGMS meter and the hexokinase method was strong.

Abstract

Background

Accurate assessment of low-level blood glucose is essential for managing neonatal hypoglycemia. This study compared low-level neonatal glucose measurements across commonly used platforms and evaluated the potential clinical implications of method-specific differences.

Methods

Paired neonatal glucose were measured using the Accu-Chek Inform II blood glucose monitoring system (BGMS) device and the Radiometer ABL90 Flex Plus blood gas analyzer. A subset of paired samples was also compared with the Beckman Coulter AU5800 hexokinase method. Method agreement was assessed using Deming regression and Bland-Altman analysis. Exploratory analyses evaluated impact of clinical characteristics, including gestational age, birth weight, growth status, and delivery mode.

Results

Agreement between Accu-Chek and ABL90 was modest (R = 0.558) with wide limits of agreement (-13.21 to 12.95 mg/dL; -0.73 to 0.72 mmol/L) and a bidirectional bias pattern. Positive-bias samples showed a mean bias of 6.13 mg/dL (0.34 mmol/L), whereas negative-bias samples showed -5.29 mg/dL (-0.29 mmol/L). By contrast, Accu-Chek demonstrated strong agreement with the AU5800 hexokinase method (R = 0.976) and narrower limits of agreement (-2.53 to 6.64 mg/dL; -0.14 to 0.37 mmol/L). No significant associations were identified between absolute bias and neonatal clinical characteristics.

Conclusions

At low glucose concentrations, agreement between the Accu-Chek and ABL90 was variable and bidirectional, while agreement between the BGMS meter and the hexokinase method was strong. The lack of association between analytical bias and neonatal characteristics suggests that platform-specific analytical performance, rather than patient factors, accounts for variability at clinically relevant glucose levels.

Read PDF

Similar papers

Open access Jan 2026

Blood Glucose Monitoring in the Neonatal Period: A Study on Evaluation Using the RAPIDPoint 500/500e.

Blood gas analyzers provide clinically useful glucose measurements in newborns, however, discrepancies from central laboratory measurements may occur under certain physiological conditions, and confirmatory testing remains essential when severe hypoglycemia is suspected.

I. Kurosawa, Hiroshi Miyabayashi, Yasue Goko et al. · 0 citations
Open access Aug 2026

Validation of Precision and Accuracy of Point-of-Care Testing Glucose Analyzers with a Standard Laboratory Method (Glucose Oxidase Method)

POCT glucose meters should be used cautiously and primarily for monitoring trends rather than definitive diagnosis or therapeutic decisions, as they have limited reliability for critical clinical decision-making.

E. Nwobodo, N. T. Emmanuel · 0 citations
Open access Sep 2026

Clinical performance and practical implications of continuous glucose monitoring in very preterm infants: a cohort study.

OBJECTIVE To explore the clinical performance of continuous glucose monitoring (CGM) in very preterm infants during the first week of life, using data from the multicentre Real time continuous glucose monitoring (REACT) randomised controlled trial (RCT) and to consider how CGM can best support neonatal glucose manageme...

G. Liu, L. Marcovecchio, K. Beardsall · 0 citations
Sep 2026

Comparison of Blood Glucose Values Used for Impending Non-Severe Hypoglycemia Management with Corresponding Continuous Glucose Monitoring Data in Type 1 Diabetes: A Secondary Analysis of the REVERSIBLE Trial.

BACKGROUND To assess continuous glucose monitoring (CGM) data corresponding to blood glucose (BG) values used for impending non-severe (NS) hypoglycemia management in people living with type 1 diabetes (PwTID). METHODS We performed a secondary analysis of the REVERSIBLE trial, a three-arm open-label crossover study a...

Andréanne Marquis, Ran Cheng, Sara Al-Mahayni et al. · 0 citations
Open access Sep 2026

ECG parameters in episodes of neonatal hyper- and hypoglycemia: an observational study

Introduction Abnormal glucose concentrations are common in the neonatal period and may entail long-term consequences. They are currently detected invasively through spot-measurements, contribute to blood loss and are painful depending on the sampling route. We hypothesized that hypo- and hyperglycemia may be associated...

L. Olivier, Julie Strunk, Camelia Lauterbach Oprea et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.