Aug 2026· Apollo Medicine· 0 citations· 20 references
TL;DR
POCT glucose measurements show excellent correlation with laboratory auto-analyser results but exhibit systematic and proportional bias, particularly at higher glucose levels, particularly at higher glucose levels.
Abstract
Point-of-care testing (POCT) for blood glucose provides rapid results and is widely used in clinical practice; however, concerns remain regarding its accuracy and agreement with laboratory auto-analyser methods. The purpose of this study was to compare glucose measurements obtained using POCT and a laboratory auto-analyser in a secondary healthcare centre within a low-resource setting.
A comparison study was conducted using 120 paired blood glucose measurements obtained simultaneously by a POCT glucometer and a laboratory auto-analyser. Descriptive statistics were used to summarise glucose values. Pearson correlation and linear regression analyses assessed the relationship between methods, while agreement was evaluated using Bland-Altman analysis. Statistical significance was set at
P
< .05.
The mean glucose concentration measured by POCT was 6.59 ± 2.27 mmol/L, while that measured by the auto-analyser was 6.33 ± 2.93 mmol/L. A strong positive correlation was observed between the two methods (
r
= 0.971,
P
< .001). Linear regression analysis yielded the equation auto-analyser = 1.249 × POCT − 1.905 (
R
² = 0.943), indicating proportional bias. Bland-Altman analysis demonstrated a mean bias of −0.26 mmol/L, with 95% limits of agreement ranging from −2.03 to +1.50 mmol/L. Greater variability in differences was observed at higher glucose concentrations.
POCT glucose measurements show excellent correlation with laboratory auto-analyser results but exhibit systematic and proportional bias, particularly at higher glucose levels. While POCT is suitable for rapid glucose assessment and monitoring, laboratory auto-analyser methods remain the preferred reference for diagnostic and critical clinical decision-making.
Both POCT devices demonstrated acceptable analytical agreement with the GOD-PAP method for blood glucose testing, however, Emtech BG Max 1 exhibited superior precision, lower analytical bias, and higher compliance with ISO 15197:2013, indicating greater reliability for routine clinical use.
Muhammad Jalil Razak, N. Ismi, Marisca Jenice Sanakya et al.· Journal Interdisciplinary He...· 0 citations
Background:Diabetes mellitus is a growing global health concern, with rising prevalence particularly in low-and middle-income countries. Accurate blood glucose estimation is essential for diagnosis, monitoring, and prevention of complications. While glucometers provide rapid, convenient results at the point ofcare, lab...
Vaaruni Sharma, Ajay K. Puri· Asian Journal of Medical Res...· 0 citations
POCT glucose meters should be used cautiously and primarily for monitoring trends rather than definitive diagnosis or therapeutic decisions, as they have limited reliability for critical clinical decision-making.
E. Nwobodo, N. T. Emmanuel· IOASD Journal of Medical and...· 0 citations
The glucometer demonstrated moderate performance and partial agreement with selected ISO 15197:2015 accuracy limits compared with laboratory-based spectrophotometric methods, but increased variability at higher glucose concentrations limits its applicability in settings requiring high analytical precision.
Pablo Letelier, Guido Fierro, K. Cuminao et al.· Journal of Medical Biochemis...· 0 citations
Blood gas analyzers provide clinically useful glucose measurements in newborns, however, discrepancies from central laboratory measurements may occur under certain physiological conditions, and confirmatory testing remains essential when severe hypoglycemia is suspected.
I. Kurosawa, Hiroshi Miyabayashi, Yasue Goko et al.· Pediatrics International· 0 citations
Reliable INR monitoring is crucial for patients on warfarin to prevent thrombotic or bleeding events. This study evaluated the accuracy of the Wondfo POCT device compared to standard laboratory methods (Stago STA-R Max3 and ACL TOP 300) in a primary healthcare setting. The goal was to assess whether this point-of-care...
Rosline Hassan, S. Iberahim, S. S. Mohd Yusoff et al.· Scientific Reports· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.