Jan 2026· Pediatrics International· Vol 68 1, pp.
e70511
· 0 citations· 15 references
Medicine
TL;DR
Blood gas analyzers provide clinically useful glucose measurements in newborns, however, discrepancies from central laboratory measurements may occur under certain physiological conditions, and confirmatory testing remains essential when severe hypoglycemia is suspected.
Abstract
Background
Rapid glucose assessment is essential in neonatal intensive care units (NICUs) for early detection of hypoglycemia. We evaluated the agreement between neonatal blood glucose levels measured using bedside blood gas analyzers (GasBS) and those measured in the central laboratory (LabBS), and identified factors associated with clinically significant discrepancies.
Methods
We retrospectively analyzed 5436 paired specimens from 2105 infants admitted to the NICU. For the primary agreement analyses, 5375 specimens with GasBS > 20 mg/dL were included. Agreement between GasBS and LabBS was assessed using Bland-Altman analysis, correlation analysis, and linear mixed models. Factors associated with clinically significant discrepancies (absolute difference ≥ 10 mg/dL) were evaluated using generalized linear mixed models. Exploratory receiver operating characteristic analysis was performed to assess the ability of GasBS to detect LabBS < 40 mg/dL.
Results
GasBS and LabBS showed good agreement, with a mean difference of 1.88 mg/dL (95% limits of agreement, -12.56 to 16.32 mg/dL) and a strong correlation (Spearman's ρ = 0.939, p < 0.001). Lower base excess, lower oxygenated hemoglobin, and higher postnatal age were associated with greater discrepancies, whereas higher hematocrit was associated with an increased risk of clinically significant discrepancies. GasBS also showed excellent performance in detecting LabBS < 40 mg/dL, with an area under the receiver operating characteristic curve of 0.985 and an optimal cutoff of 43.5 mg/dL.
Conclusions
Blood gas analyzers provide clinically useful glucose measurements in newborns. However, discrepancies from central laboratory measurements may occur under certain physiological conditions, and confirmatory testing remains essential when severe hypoglycemia is suspected.
At low glucose concentrations, agreement between the Accu-Chek and ABL90 was variable and bidirectional, while agreement between the BGMS meter and the hexokinase method was strong.
Alyssa M. Lee, Li-Yun Cao· The Journal of Applied Labor...· 0 citations
POCT glucose measurements show excellent correlation with laboratory auto-analyser results but exhibit systematic and proportional bias, particularly at higher glucose levels, particularly at higher glucose levels.
Ufuoma Ohwo, K. Digban, C. Okafor· Apollo Medicine· 0 citations
Background:Diabetes mellitus is a growing global health concern, with rising prevalence particularly in low-and middle-income countries. Accurate blood glucose estimation is essential for diagnosis, monitoring, and prevention of complications. While glucometers provide rapid, convenient results at the point ofcare, lab...
Vaaruni Sharma, Ajay K. Puri· Asian Journal of Medical Res...· 0 citations
The glucometer demonstrated moderate performance and partial agreement with selected ISO 15197:2015 accuracy limits compared with laboratory-based spectrophotometric methods, but increased variability at higher glucose concentrations limits its applicability in settings requiring high analytical precision.
Pablo Letelier, Guido Fierro, K. Cuminao et al.· Journal of Medical Biochemis...· 0 citations
ABG potassium estimation is a reliable alternative for rapid assessment in emergency, life-threatening situations but should not replace laboratory testing, while ABG potassium may be usable for rapid bedside assessment.
Poulami Samanta, Somnath Das, Rana Barik et al.· International Journal of Hea...· 0 citations
INTRODUCTION
Glycemic management in the intensive care unit (ICU) remains challenging because glucose levels fluctuate rapidly and monitoring is intermittent. Interstitial continuous glucose monitoring (CGM) could improve surveillance, but its performance during rapid glucose excursions remains uncertain. We assessed t...
M. Seret, C. Dupuis, Vanessa Alves de Araújo et al.· Journal of Diabetes Science...· 0 citations