Are our self-care devices accurate and on the mark? Validation of a diabetes point-of-care device

Aug 2026 · MRIMS journal of health sciences · 0 citations

TL;DR

A diagnostic accuracy cross-sectional study was conducted among patients with diabetes mellitus type 2 in the urban field practice area of a medical college, among 240 patients, which showed that POCD has excellent reliability.

Abstract

A point-of-care device (POCD) is a medical diagnostic tool designed for use at or near the patient, providing rapid results outside traditional laboratory settings. The term “Gold Standard” refers to a diagnostic tool that is considered the accurate benchmark for diagnosis. A glucometer is a popularly used point-of-care device for the diagnosis of diabetes, which, when validated against the gold standard, saves time, money, and manpower resources invested into the diagnosis of diabetes. To validate a POCD, i.e., glucometer, against the gold standard method (hexokinase method). A diagnostic accuracy cross-sectional study was conducted from September 2025 to December 2025 among patients with diabetes mellitus type 2 in the urban field practice area of a medical college, among 240 patients. Blood glucose levels were measured using the POCD glucometer and the gold standard method, i.e. the hexokinase method, simultaneously. Data entry and analysis were done. Validity testing of the POCD was performed. Sensitivity and specificity were found to be 84.4% and 71.1%, respectively. Positive predictive value (80.2%), negative predictive value (76.6%), and accuracy (78.8%), all confirmed the validity of POCD. Area under the curve (AUC) =0.932 showed the good performance of the POCD. The Bland–Altman plot showed the good concordance of the POCD with the gold standard. Intraclass correlation coefficient = 0.985, which showed that POCD has excellent reliability. The glucometer used in the study is an efficient tool to use in mass screening programs.

Read PDF

Similar papers

Open access Aug 2026

Validation of Precision and Accuracy of Point-of-Care Testing Glucose Analyzers with a Standard Laboratory Method (Glucose Oxidase Method)

POCT glucose meters should be used cautiously and primarily for monitoring trends rather than definitive diagnosis or therapeutic decisions, as they have limited reliability for critical clinical decision-making.

E. Nwobodo, N. T. Emmanuel · 0 citations
Open access Aug 2026

Validation of point-of-care INR testing against standard laboratory methods in a Malaysian primary healthcare setting

Reliable INR monitoring is crucial for patients on warfarin to prevent thrombotic or bleeding events. This study evaluated the accuracy of the Wondfo POCT device compared to standard laboratory methods (Stago STA-R Max3 and ACL TOP 300) in a primary healthcare setting. The goal was to assess whether this point-of-care...

Rosline Hassan, S. Iberahim, S. S. Mohd Yusoff et al. · 0 citations
Open access Aug 2026

Assessing the reliability and validity for metabolic syndrome diagnosis using point-of-care analyzers in a mobile setup - a pilot study

The metabolic syndrome (MetS) is characterized by having three or more of these risk factors: increased blood glucose (Glc), serum triglycerides (TG), blood pressure (BP), waist circumference (WC), and reduced high-density lipoprotein (HDL). Mobile diagnostics using point-of-care (POC) blood analyzers could help identi...

S. Houtenbos, A. Schraplau, P. Wippert et al. · 0 citations
Review Open access Sep 2026

Accuracy of Point-of-Care Blood Glucose Meters Compared With Central Laboratory Reference Methods in Critically Ill AdultsA Systematic Review

Background: Point-of-care (POC) blood glucose meters shorten turnaround time for bedside glycemic management, but critical illness introduces physiologic and analytical conditions that can degrade agreement with central laboratory measurements. Objective: To synthesize evidence on the analytical and clinical accuracy o...

C. Amorado, Leah Quinto, Mark Kerwin F. Sayas et al. · 1 citation
Oct 2026

A-362 Analytical validation of a semi-automated portable IoT-enabled POCT device for screening of non-communicable diseases

Mobilab Duo is a semi-automated, IoT-enabled, battery-operated, point-of-care testing (POCT) device designed in the form of a compact, portable suitcase (Figure 1) developed by Primary Health Tech Pvt. Ltd. in collaboration with IIT Guwahati, India, for screening of non-communicable diseases. This device is desig...

Saumya Srivastava, S. Datta, Milan Bhandari et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.