Fast ungated five-dimensional cardiac MRI on a 1.5 T MR-linac for MRI-guided radiotherapy

Jul 2026 · 0 citations · 44 references
Physics

TL;DR

A fast, ungated 5D-MRI reconstruction method based on the CMR-MOTUS framework enables rapid, high-quality, and personalized motion characterization, demonstrating potential for integration into MRI-guided STAR treatments.

Abstract

Background: Stereotactic arrhythmia radio-ablation (STAR) for patients with ventricular tachycardia is currently limited by complex cardiorespiratory motion. Current 5D-MRI motion models require long acquisition and reconstruction times, limiting clinical viability. Objective: To develop a fast, ungated 5D-MRI reconstruction method for personalized motion characterization to support MRI-guided STAR treatments. Methods: We propose a fast, ungated 5D-MRI reconstruction method based on the CMR-MOTUS framework. The method uses a 3D Cartesian acquisition with a joint optimization framework to reconstruct a motion-corrected reference image and low-rank deformation vector fields (DVFs). By exploiting the low rank structure, we explicitly disentangle respiratory and cardiac motion during optimization. Then, the DVFs are used for 5D-MRI reconstruction with a retrospectively adjustable number of motion states. Validation was performed using digital and physical cardiorespiratory phantoms. Furthermore, the approach was evaluated using 10 healthy volunteers, comparing motion consistency with 2D cine MRI. Results: Validation of 5D CMR-MOTUS using digital and physical phantoms demonstrated accurate 5D-MRI reconstruction. In the physical phantom, 5D CMR-MOTUS achieved a left-ventricle DICE of 0.96 +/- 0.01. In the volunteer cohort, the 5D-MRI scans showed strong motion to 2D cine MRI, with a cardiac motion error of 0.1 +/- 0.9 mm and a respiratory motion error of 0.2 +/- 2.9 mm. Crucially, 5D-MRI data were acquired in 1 minute and reconstructed in 6 minutes. Conclusions: The proposed 5D-MRI method enables rapid, high-quality, and personalized motion characterization, demonstrating potential for integration into MRI-guided STAR treatments. Data Availability: The 3D k-space data and 5D reconstructions for the ten volunteers are publicly available at https://doi.org/10.5281/zenodo.21278894

View source

Similar papers

Open access Aug 2026

Motion-Robust Approach for 4D Cardiac Cine Using 2D Real-Time Acquisitions and Slice-to-Volume Reconstruction.

The RT-SVR approach enables 4D cardiac cine with comparable bi-ventricular volumes, contrast, sharpness, and acquisition time compared to standard breath-hold cine while providing additional motion robustness.

Ye Tian, A. A. Joshi, Jon A. Detterich et al. · 0 citations
Open access Aug 2026

Volumetric real-time MRI for the guidance of cardiac catheterization at 0.55T.

Volumetric real-time MRI is feasible for the guidance of invasive procedures such as right-heart catheterization at 0.55T and offers flexible real-time re-slicing, volumetric 3D visualization, and the potential for improved device monitoring.

Prakash Kumar, R. Ramasawmy, A. Javed et al. · 0 citations
Open access Sep 2026

Towards Faster Fetal 3D CINE Cardiac MRI Reconstruction Using Region-Optimized Virtual Coils (ROVir).

A computationally efficient image-reconstruction pipeline for 3D balanced steady state free precession CINE fetal cardiac MRI that reduces radial streaking artifacts, increases effective temporal resolution, and improves diagnostic image quality compared with previous work is developed and evaluated.

M. Piek, J. Töger, E. Heiberg et al. · 0 citations
Open access Oct 2026

Non-Rigid Motion-Compensated MR Multitasking for Free-Breathing Low-Dose Dynamic Contrast-Enhanced MRI in the Abdomen.

PURPOSE To improve free-breathing low-dose quantitative dynamic contrast-enhanced (DCE) MRI for imaging pancreatic ductal adenocarcinoma. METHODS We introduced non-rigid (NR) motion compensation (MoCo) into MR Multitasking using a novel "in-subspace" formulation that directly applies motion fields to spatial feature...

Ling-Ceng Ma, Chao-Ying Wu, Li-Xia Wang et al. · 0 citations
Sep 2026

High-dose-rate prostate brachytherapy imaging optimization: A motorized sagittal three-dimensional ultrasound acquisition system.

The motorized attachment for SR3D acquisition reduced image variability and improved image quality, and may encourage broader use of sagittal reconstruction in HDR-BT, potentially improving patient outcomes.

David Contella, Chandima Edirisinghe, Kevin Barker et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.