Aug 2026· Journal Interdisciplinary Health· Vol 2, pp. 113-123· 0 citations· 23 references
TL;DR
Both POCT devices demonstrated acceptable analytical agreement with the GOD-PAP method for blood glucose testing, however, Emtech BG Max 1 exhibited superior precision, lower analytical bias, and higher compliance with ISO 15197:2013, indicating greater reliability for routine clinical use.
Abstract
Introduction: Point-of-care testing (POCT) for blood glucose measurement is increasingly used because it provides rapid results for clinical decision-making. However, its analytical performance must be evaluated before routine implementation. This study aimed to compare the precision, accuracy, and agreement of the Easy Touch and Emtech BG Max 1 POCT devices with the Glucose Oxidase–Peroxidase (GOD-PAP) reference method for blood glucose measurement.
Research Methodology: An observational analytic study with a cross-sectional design was conducted at the Buton Tengah Regional General Hospital Laboratory, Indonesia. Fifty venous blood samples were analyzed using the Easy Touch and Emtech BG Max 1 POCT devices and the GOD-PAP laboratory method as the reference standard. Precision was evaluated using the coefficient of variation (CV), accuracy was assessed through percentage bias and compliance with ISO 15197:2013, and agreement was determined using nonparametric Bland–Altman analysis.
Results: The CV values were 5.54% for Easy Touch, 1.76% for Emtech BG Max 1, and 0.80% for the GOD-PAP method. The mean bias was 7.74% for Easy Touch and −5.78% for Emtech BG Max 1. Compliance with ISO 15197:2013 was achieved in 62% of Easy Touch measurements and 90% of Emtech BG Max 1 measurements. Bland–Altman analysis demonstrated good agreement between both POCT devices and the reference method. However, Emtech BG Max 1 showed narrower limits of agreement and closer agreement with GOD-PAP than Easy Touch.
Conclusion: Both POCT devices demonstrated acceptable analytical agreement with the GOD-PAP method for blood glucose testing. However, Emtech BG Max 1 exhibited superior precision, lower analytical bias, and higher compliance with ISO 15197:2013, indicating greater reliability for routine clinical use. Periodic analytical performance evaluation is recommended to maintain the quality and accuracy of POCT-based blood glucose testing.
POCT glucose measurements show excellent correlation with laboratory auto-analyser results but exhibit systematic and proportional bias, particularly at higher glucose levels, particularly at higher glucose levels.
Ufuoma Ohwo, K. Digban, C. Okafor· Apollo Medicine· 0 citations
POCT glucose meters should be used cautiously and primarily for monitoring trends rather than definitive diagnosis or therapeutic decisions, as they have limited reliability for critical clinical decision-making.
E. Nwobodo, N. T. Emmanuel· IOASD Journal of Medical and...· 0 citations
Background: Point-of-care testing (POCT) devices offer a rapid alternative for lipid assessment, but comprehensive analytical validation against reference laboratory methods remains limited. Objective: This study aimed to evaluate the agreement, systematic bias, and diagnostic performance of the Mission® Cholesterol 3-...
S. Guzmán-Aquino, Roberto Rivera-Sánchez, K. A. Aguayo-Cerón et al.· Diagnostics· 0 citations
Background: Point-of-care (POC) blood glucose meters shorten turnaround time for bedside glycemic management, but critical illness introduces physiologic and analytical conditions that can degrade agreement with central laboratory measurements. Objective: To synthesize evidence on the analytical and clinical accuracy o...
C. Amorado, Leah Quinto, Mark Kerwin F. Sayas et al.· International Journal of Hea...· 1 citation
A diagnostic accuracy cross-sectional study was conducted among patients with diabetes mellitus type 2 in the urban field practice area of a medical college, among 240 patients, which showed that POCD has excellent reliability.
Karthikeyan Jayakumar, Misha Gorantla· MRIMS journal of health scie...· 0 citations
Point-of-care testing (POCT) for glucose monitoring plays a crucial role in enabling timely clinical decisions within healthcare settings. The Roche cobas® pulse is a POCT device for glucose measurement and is compatible with cobas® GLU test strips and cobas® GLU QC kit for internal quality control (Roche Diagnos...
Pit Hung Lim, S. Lim, Kelly Lim et al.· Clinical Chemistry· 0 citations