2026· Journal of Medical Biochemistry· Vol 45, pp. 1999-2008· 0 citations· 46 references
TL;DR
The glucometer demonstrated moderate performance and partial agreement with selected ISO 15197:2015 accuracy limits compared with laboratory-based spectrophotometric methods, but increased variability at higher glucose concentrations limits its applicability in settings requiring high analytical precision.
Abstract
Background: The accuracy of portable blood glucose
monitoring systems is critical for the safe management
and follow-up of Diabetes Mellitus. This study evaluated the accuracy of a capillary glucose monitoring device
compared with laboratory-based spectrophotometric
analysis of venous blood according to ADA recommendations and selected ISO 15197:2015 accuracy limits.
Methods: A cross-sectional study was conducted using
133 paired venous and capillary blood samples. Glucose
concentrations were measured using the Humalyzer
2000 and Mindray BA-88A spectrophotometers and the
Prodigy AutoCode® glucometer. Data were analysed using Kruskal-Wallis and Spearman correlation tests, with
p<0.05 considered statistically significant.
Results: The glucometer reported slightly lower glucose
values than the laboratory-based spectrophotometric
methods. Correlation showed a weak association with the
Humalyzer 2000 (r=0.17) and a moderate association
with the Mindray BA-88A (r=0.415). Seventy-four per
cent of measurements were within the ISO 15197:2015
acceptance limits compared with the Humalyzer 2000,
whereas 92% were within the acceptance limits compared with the Mindray BA-88A for glucose concentrations <100 mg/dL. Bland–Altman analysis showed low
systematic bias versus both analysers (-0.44 to -3.89 mg/
dL), although agreement was more consistent with the
Mindray BA-88A because of narrower limits of agree ment. Greater variability was observed at higher glucose
concentrations, particularly with the Humalyzer 2000.
Conclusion: The glucometer demonstrated moderate
performance and partial agreement with selected ISO
15197:2015 accuracy limits compared with laboratory-based spectrophotometric methods. Increased variability at higher glucose concentrations limits its applicability
in settings requiring high analytical precision; therefore,
results should be interpreted with caution.
POCT glucose meters should be used cautiously and primarily for monitoring trends rather than definitive diagnosis or therapeutic decisions, as they have limited reliability for critical clinical decision-making.
E. Nwobodo, N. T. Emmanuel· IOASD Journal of Medical and...· 0 citations
POCT glucose measurements show excellent correlation with laboratory auto-analyser results but exhibit systematic and proportional bias, particularly at higher glucose levels, particularly at higher glucose levels.
Ufuoma Ohwo, K. Digban, C. Okafor· Apollo Medicine· 0 citations
Objective: The aim of this study was to retrospectively evaluate the performance of 2 glucometer devices using the accuracy criteria specified in ISO 15197:2015.
Methods: Glucose results obtained from Accu-Chek Inform II and Sejoy BG-707 glucometers, which measure according to the glucose dehydrogenase principle, were...
T. Noyan, Murat Cihan, E. Yiğit et al.· Journal of Turkish Clinical...· 0 citations
At low glucose concentrations, agreement between the Accu-Chek and ABL90 was variable and bidirectional, while agreement between the BGMS meter and the hexokinase method was strong.
Alyssa M. Lee, Li-Yun Cao· The Journal of Applied Labor...· 0 citations
HbA1c is an approved test for the diagnosis and monitoring of diabetes. Several assays are in use worldwide based on chromatography, cation-exchange, enzymatic and immunological principles. High Performance Liquid Chromatography (HPLC) is regarded as the gold standard method while other assays including enzymatic assay...
Introduction: the use of "point-of-care" (POC) monitoring, like blood gas analysis, allows clinicians to measure important variables in real-time, allowing for immediate results to be provided to make rapid decisions. However, the reliability of basic biochemistry and hematology parameters in arterial blood conducted a...
Maaroufi Ayoub, Diai Abdellatif, El Akef Taoufik et al.· Primary Health Care Practice...· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.